Isolating Converter Typ: KFD0 SCS

application: Digital two-wire transmitter converts Pt-100, DIN EN 60751 measuring resistors
into a 4…20mA current signal.
For installation on a standard rail according to EN50022 (DIN, top hat rail).

Get offer
Case
Plastics
Case dimensions
B x H x T : 13 x 110 x 110 mm
Protection class
Case: IP68 Clamps: IP 00
electrical connection
Clamps, Cable cross section: till 1 x 1,5 mm²
adjustment range
-200°C...0...+850°C
Burden
Rb [ Ohm ] = ( Ub [ V ] - U-Auxiliary energy_min [ V ] ) / 0,0215 [ A ]
Response time
< 2 s
Ambient temperatures
- 40 ... + 85°C
Weight
ca. 0,40 kg
Ex Schutz
II 1 G Ex ia IIC T6/T5/T4 Ga II 1 D Ex ia IIIC T135°C/T100°C/T85°C Da II 3 G Ex nA IIC T6/T5/T4 Gc
Option
see technical data sheet