Pressure transmitters with thick film strain gauge, ceramic, ATEX Typ: CS

from 0-1 bar to 0-400 bar
Accuracy class 0,3 FS

 

Service intended: Relative and absolute pressure measurement of gaseous and liquid media, the flush diagram type also for highly viscous and crystallizing pressure media.

Fields of application:
● Plant and mechanical engineering
● ventilation and air-conditioning technology
● Pneumatics, hydraulics

Get offer
Case
Stainless steel 304S
Degree of protection
IP 65 per EN 60529 / IEC 529
Pressure connection
Ceramic diaphragm, Measurig bridge, thick film strain gauge
Sensing element
Ceramic diaphragm, Measurig bridge, thick film strain gauge
Pressure and overpressure ranges
see Technical Details page 05.0302
Electrical connection
L-plug EN175301-803 Form A
Signal output Power supply
4 ... 20 mA, 2-wire Ub = 12 ... 30 V DC Load: Rb [Ohm] : (Ub [ V ]-12 [ V ] )/0,025 [ A ]
Intrinsic safety- BVS 07 ATEX E 177
II 2G Ex ia IIC T4-T6 Gb II 2D Ex ia IIIC T135°C Db
Bleeder resistor
RAb >= 1 MOhm
Response time
t < 10 ms
Power consumption
P max = 0,75 W
EMC-Directive
2004/108/EC EN 61326 Emission (Group 1, Class B. Immunity, industrial locations)
Permissible temperatures Tmin / Tmax
Storage temperature - 40 ... +100°C Ambient temperature - 20 ... + 70°C II 2G Ex ia IIC T4 Gb - 20 ... + 70°C II 2G Ex ia IIC T6 Gb - 20 ... + 40°C II 2D Ex ia IIIC T135°C Db - 20 ... + 70°C Medium temperature - 25 ... + 80°C Compensated range - 10 ... + 80°C
Accuracy
Linearity + hysteresis + repeatability <0,3% FS / 0,2 % BFSL
Temperature error
Mean TK of zero < 0,2 % FS / 10 K Mean TK of range < 0,2 % FS / 10 K
Weight
0,27 kg
Optional extras
1) see digital indicator
  • product data

  • customer data

  • This field is for validation purposes and should be left unchanged.

Diese Seite benutzt Cookies. mehr Info
OK
Skip to toolbar