Pressure transmitters with piezoresistive sensing principle, membrane stainless steel, ATEX Typ: PS

from 0-40 m to bis 0-16 bar
Accuracy class 0,3 FS

 

Service intended: Relative and absolute pressure measurement of low and medium pressure ranges, gaseous and liquid media, the flush diagram type also for highly viscous and crystallizing pressure media.

Fields of application:
● Plant and mechanical engineering
● ventilation and air-conditioning technology

Get offer
Case
Stainless steel 304S
Degree of protection
IP 65 per EN 60529 / IEC 529
Pressure connection
Stainless steel 316 Ti G1/2B per EN 837-1 G3/4B with flush diaphragm G1/4 female
Sensing element
st. st. 316L, Measuring bridge: piezoresistive
Pressure and overpressure ranges
see Technical Details page 05.0102
Electrical connection
L-plug EN175301-803 Form A
Signal output Power supply
4 ... 20 mA, 2-wire Ub = 12 ... 30 V DC 0 ... 20 mA, 3-wire Load:Rb [Ohm] : (Ub [ V ]-12 [ V ] )/0,025 [ A ]
Intrinsic safety- BVS 07 ATEX E 177
II 2G Ex ia IIC T4-T6 Gb II 2D Ex ia IIIC T135°C Db
Bleeder resistor
RAb >= 1 MOhm
Response time
t < 10 ms
Power consumption
P max = 0,75 W
EMC-Directive
2004/108/EC EN 61326 Emission (Group 1, Class B. Immunity, industrial locations)
Permissible temperatures Tmin / Tmax
Storage temperature - 40 ... +100°C Ambient temperature - 20 ... + 70°C II 2G Ex ia IIC T4 Gb - 20 ... + 70°C II 2G Ex ia IIC T6 Gb - 20 ... + 40°C II 2D Ex ia IIIC T135°C Db - 20 ... + 70°C Medium temperature - 25 ... + 80°C Compensated range - 10 ... + 80°C
Accuracy
Linearity + hysteresis + repeatability <0,3% FS / 0,2 % BFSL
Temperature error
Mean TK of zero < 0,2 % FS / 10 K Mean TK of range < 0,2 % FS / 10 K
Weight
0,27 kg
Optional extras